 Model #: Ti25
 Fluke Ti25 Thermal Imager Camera Gun
 Fluke Ti25 Thermal Imager Camera Gun    
Ti25 Thermal Imager

Features:

  • Enhanced problem detection and analysis capabilities with IR-Fusion® Technology. Simply scroll through the different viewing modes quickly to better identify trouble areas in Full IR thermal, picture-in-picture, or automatic blend visual and thermal images.
  • Optimized for field use in harsh work environments.
  • Engineered and tested to withstand a 2 m (6.5 ft) drop—When was the last time you dropped a tool?
  • Withstands dust and water—tested to an IP54 rating.
  • Delivers the clear, crisp images needed to find problems fast.
  • Identify even small temperature differences that could indicate problems with excellent thermal sensitivity (NETD).
  • Intuitive, three-button menu is easy to use—simply navigate with the push of a thumb.
  • No need to carry pen and paper—record findings by speaking into the camera. Voice annotations can be recorded with every image you take. Voice comments are saved along with individual images for future reference (Ti25 only).
  • Everything needed to get started is included.
  • Adjustable hand strap for left-or right-handed use.
  • Manufactured in the U.S.A.

Detailed specifications
Temperature
Temperature measurement range (not calibrated below -10 °C) -20 °C to +350 °C (-4 °F to + 662 °F) (two ranges)
Accuracy ± 2 °C or 2 % (whichever is greater)
On-screen emissivity correction Yes
Imaging performance
Field of view 23° x 17°
Spatial resolution (IFOV) 2.5 mRad
Minimum focus distance Thermal lens: 15 cm (6 in)
Visible (visual) light lens: 46 cm (18 in)
Focus Manual
Image frequency 9 Hz refresh rate
Detector type 160 X 120 focal plane array, uncooled microbolometer
Infrared lens type 20 mm F = 0.8 lens
Thermal sensitivity (NETD) ≤ 0.1 °C at 30 °C (100 mK)
Infrared spectral band 7.5 μm to 14 μm
Visual camera 640 x 480 resolution
Image presentation
Palettes Ironbow, blue-red, high contrast, amber, hot metal, grey
Level and span Smooth auto-scaling and manual scaling of level and span
Minimum span (in manual mode) 2.5 °C (4.5 °F)
Minimum span (in auto mode) 5 °C (9 °F)
IR-Fusion® information Full infrared with MAX, MID, or MIN automatic blending; and picture-in-picture with MAX, MID, or MIN automatic blending (Visual and IR blending)
Picture-In-Picture (PIP) Three levels of on-screen IR blending displayed in center 320 x 240 pixels
Full screen (PIP off) Three levels of on-screen IR blending displayed in center 640 x 480 LCD
Voice annotation
Voice annotation 60 seconds maximum recording time per image
Image and data storage
Storage medium SD memory card (2 GB memory card will store at least 1200 fully radiometric (.is2) IR and linked visual images each with 60 seconds voice annotations or 3000 basic (.bmp) IR images)
File formats Non-radiometric (.bmp) or fully-radiometric (.is2)
No analysis software required for non-radiometric (.bmp) files
Export file formats w/SmartView™ software JPEG, BMP, GIF, PNG, TIFF, WMF, EXIF, and EMF

General specifications
Temperature
Operating: -10 °C to 50 °C (14 °F to 122 °F)
Storage: -20 °C to +50 °C (-4 °F to 122 °F) without batteries
Relative humidity
10 % to 90 % non-condensing
Display
9.1 cm (3.6 in) diagonal landscape color VGA (640 x 480) LCD with backlight (selectable bright or auto)
Controls and adjustments
User selectable temperature scale (°C/°F)
Language selection
Time/Date set
Software
SmartView™ full analysis and reporting software included
Power
Battery: Internal rechargeable battery pack (included)
Battery life: Three to four hours continuous use (assumes 50 % brightness of LCD)
Battery charge time using ac adapter/charger and dc car charger: Two hours for full charge
AC operation/charging
AC adapter/charger (110 Vac to 220 Vac, 50 Hz to 60 Hz). Charges battery while imager is operating. Universal ac mains adapters included.
Power saving
Sleep mode activated after 5 minutes of inactivity, automatic power off after 20 minutes of inactivity
Safety standards
CE Directive: IEC/EN 61010-1 2nd Edition Pollution Degree 2
Electromagnetic compatibility
EMC directive: EN61326-1
C-Tick: IEC/EN 61326
US FCC: CFR 47, Part 15 Class A
Vibration
2 G, IEC 68-2-29
Shock
25 G, IEC 68-2-29 {2 m (6.5 ft) drop, 5 sides}
Dimensions (HxWxL)
0.27 m x 0.13 m x 0.15 m (10.5 in x 5 in x 6 in)
Weight
1.2 kg (2.65 lb)
IP rating
IP54
Warranty
Two-years
Calibration cycle
Two-years (assumes normal operation and normal aging)
Supported languages
English, Italian, German, Spanish, French, Russian, Portuguese, Swedish, Turkish, Czech, Polish, Finnish, Simplified Chinese, Traditional Chinese, Korean, and Japanese